User Tools

Site Tools


macroct

This is an old revision of the document!


macroCT

X-Ray Source

PSX10-WB-130 Microfocus Source

  • 20kV-130kV targe voltage
  • 0.356mA@45kV 0.123mA@130kV beam current
  • 7µm@4W - 21µm@16W focal spot on tungsten target
  • 115° cone angle
  • 0.51mm Beryllium window

On /dev/ttyS3 with 3840bps 8N1

Detectors

Dexis Platinum / GXS-700

  • SN: 0104430711
  • 19.5µm x 19.5µm Pixels
  • 1324×1692 Resolution
  • Sensor Firmware Version: 0.5.10, 0.3.6, 0.4.5
  • CPLD Version: 3.3.0.0
  • API Version: 1.0.10.0
  • VID 5328 PID 2010
  • SN: 1093330312
  • 19.5µm x 19.5µm Pixels
  • 1552×1040 Resolution

https://nucwiki.org/wiki/index.php/GXS700_FOSS

https://nucwiki.org/wiki/index.php/Gendex_GXS700

https://nucwiki.org/wiki/index.php/Gendex_GXS700_Dump

https://github.com/JohnDMcMaster/gxs700

Dexela

paxScan

I/O

PWM

Channels

  • 0: Chamber Light
  • 1: Alignment Laser Vertical
  • 2: Alignment Laser Horizontal
  • 3: (unused)
  • 4: Case Fan 1
  • 5-15: (unused)

Outputs

Channels:

  • 0: Stepper Motor Supply
  • 1: (unused)
  • 2: (unused)
  • 3: (unused)
  • 4: (unused)
  • 5: (unused)
  • 6: (unused)
  • 7: (unused)
  • 8: Tube Supply
  • 9: (unused)
  • 10: Software E-Stop
  • 11: Safety Start
  • 12: Software Power Interlock
  • 13: Software Chamber Interlock
  • 14: 15V Detector Supply
  • 15: 24V Detector Supply

Mechanics

Stepper Motors

GRBL

Settings
$0=10
$1=254
$2=0
$3=0
$4=0
$5=1
$6=0
$10=1
$11=0.020
$12=0.002
$13=0
$20=0
$21=1
$22=1
$23=0
$24=100.000
$25=500.000
$26=250
$27=5.000
$30=12000
$31=550
$32=0
$100=160.000
$101=160.000
$102=160.000
$103=160.000
$110=600.000
$111=600.000
$112=600.000
$113=600.000
$120=40.000
$121=40.000
$122=40.000
$123=40.000
$130=400.000
$131=400.000
$132=400.000
$133=400.000