| - | ===== I/O ===== | + | The USB connector is actually Ethernet. |
| + | d+/d- and gnd/ID are the ethernet lines. | ||
| + | If you hook up to that, you can see a autoconf IP address. | ||
| + | Access the webinterface and login with admin: | ||
| + | This URL allows you a bind shell: http:// | ||
| + | Run this: | ||
| + | mount -o remount,rw /home | ||
| + | echo " | ||
| + | echo "sed -i ' | ||
| + | echo "sed -i ' | ||
| + | sync | ||
| + | mount -o remount,ro /home | ||
| + | |||
| + | This allows you to login as vsp with password password and then becoming root with " | ||
| + | |||
| + | ssh: vsp: | ||
| + | su: root: | ||
| + | |||
| + | IP: 192.168.179.200 | ||
| + | ===== I/O ===== | ||
| + | on / | ||
| ==== PWM ==== | ==== PWM ==== | ||
| Channels | Channels | ||
| Line 126: | Line 146: | ||
| $133=400.000 | $133=400.000 | ||
| </ | </ | ||
| + | |||
| + | ==== Startup Sequence ==== | ||
| + | |||
| + | Connect to / | ||
| + | |||
| + | Enable Debug Mode | ||
| + | / | ||
| + | |||
| + | Enable Sensor Supply | ||
| + | / | ||
| + | |||
| + | Enable Tube Supply | ||
| + | / | ||
| + | |||
| + | Connect to /dev/ttyS3 | ||
| + | |||
| + | Check Tube Status | ||
| + | /dev/ttyS3 : status | ||
| + | |||
| + | Check Beam Voltage HV | ||
| + | /dev/ttyS3 : hv setting | ||
| + | |||
| + | Check Beam Current | ||
| + | /dev/ttyS3 : beam setting | ||
| + | |||
| + | Set Software Chamber Interlock | ||
| + | / | ||
| + | |||
| + | Set Software Power Interlock | ||
| + | / | ||
| + | |||
| + | Turn Key-Switch to ON | ||
| + | |||
| + | Release Emergency Stop | ||
| + | |||
| + | Release Software E-Stop | ||
| + | / | ||
| + | |||
| + | Pulse Software Safety Start | ||
| + | / | ||
| + | / | ||
| + | |||
| + | Check Tube Status (should be "Safe Infocus", | ||
| + | /dev/ttyS3 : status | ||
| + | |||
| + | Override Tube Conditioning | ||
| + | /dev/ttyS3 : program 10 | ||
| + | |||
| + | Start Sensor Capture | ||
| + | cd ~/ | ||
| + | python3 test.py | ||
| + | |||
| + | Enable XRAYs | ||
| + | /dev/ttyS3 : xray on | ||
| + | |||
| + | Check Tube Status to see when current has ramped up | ||
| + | /dev/ttyS3 : status | ||
| + | /dev/ttyS3 : status | ||
| + | ... | ||
| + | |||
| + | Trigger Sensor Exposure | ||
| + | |||
| + | Disable XRAYs | ||
| + | /dev/ttyS3 : xray off | ||
| + | |||
| + | Turn off Key-Switch | ||
| + | |||
| + | Turn off safety | ||
| + | / | ||
| + | / | ||
| + | / | ||
| + | |||
| + | Turn off Tube supply | ||
| + | / | ||
| + | |||
| + | Turn off Sensor supply | ||
| + | / | ||
| + | |||
| + | Disconnect Serial ports | ||
| + | |||
| + | Shutdown PC | ||