User Tools

Site Tools


microct

This is an old revision of the document!


µCT

X-Ray Source

PXS5-724EA Microfocus Source

  • 20kV-70kV targe voltage
  • 0-0.1 mA beam current
  • 10µm x 10µm focal spot on tungsten target
  • 34° (or 40°) cone angle
  • 0.13mm Beryllium windows

Connector

  1. Chassis Ground
  2. 4.0V DC kV Control Source
  3. HV Programming (4.0V = 80kV)
  4. HV Programming Ground
  5. Current Programming (2.5V = 100µA)
  6. Standby Power (+12V DC)
  7. HV Monitor Voltage (4.0V = 80kV)
  8. Current Monitor (2.5V = 100µA)
  9. Monitor Ground
  10. +2.5V DC mA Control Source
  11. +12 DC Operate Power
  12. DC Operate Ground

Mating Connector :

Thoshiba DGO 73

  • 65kV
  • 8mA beam current
  • 700µm x 700µm focal spot
  • 2mm Aluminium filter (integrated in tube?)

X-Ray Detector

Dexis Platinum / GXS-700

  • SN: 0104430711
  • 19.5µm x 19.5µm Pixels
  • 1324×1692 Resolution
  • Sensor Firmware Version: 0.5.10, 0.3.6, 0.4.5
  • CPLD Version: 3.3.0.0
  • API Version: 1.0.10.0
  • VID 5328 PID 2010
  • SN: 1093330312
  • 19.5µm x 19.5µm Pixels
  • 1552×1040 Resolution

https://nucwiki.org/wiki/index.php/GXS700_FOSS https://nucwiki.org/wiki/index.php/Gendex_GXS700 https://nucwiki.org/wiki/index.php/Gendex_GXS700_Dump https://github.com/JohnDMcMaster/gxs700

DigiPan

  • 1244 x 61 pixels
  • 104µm pixel pitch
  • 12bits per pixel
  • 1 Start bit, 1 Control bit (always 1)

https://hackaday.io/project/62710-taking-digital-xray-shots-for-cheap-300

Connector

  1. (BL) Ground
  2. (RD) +5V
  3. (OR) +15V
  4. (GR) -15V
  5. (BL) RS422 A
  6. (WH) RS422 B

Flatpanel Detector

Fluorescent Screen

Mechanics

Rotary Stage

  • 0.1° minimum resolution
  • AMS AS5147 Magnetic Encoder (On Axis) Datasheet
  • Wormgear drive (AliExpress
    • 29.5mm Center Distance
    • ~50mm gear diameter

Source & Detector

  • Distances must be known
    • Source - Center of Rotation (SOD - Source Object Distance)
    • Center of Rotation - Detector (ODD - Object Detector Distance)
    • Source - Detector (SDD - Source Detector Distance)

Reconstruction

Correction

Convert x-ray images to projections:

p(s) = -ln(I(s) / I(0))

p(s) : Projected Intensity I(s) : Intensity recorded at detector after passing through an object I(0) : Intensity at source (white reference)

Filtered Backprojection